GAW Precipitation Chemistry Manual –Laboratory Operations- Conductivity (27 December 2018)
6
5) Prepare samples for analysis. Make sure that each tube has a minimum volume. Minimum
volumes will vary according to injection loop size and loop rinse. Cover each tube opening with
Parafilm® or with a cap that can be pierced. Place the tubes in order in the sample changer rack.
6) Check for a stable pump pressure and conductivity.
7) Check the DI water chromatograms for the correct shape. The shape of the chromatogram
depends on the eluent type. (a) When a carbonate eluent is used, the chromatogram will have a
water dip usually two to three minutes into the chromatogram. The water dip occurs right before
the chloride peak. See Figure 4.23 for the DI water chromatogram using carbonate eluent. Note
that the scale on the Y-axis is -4.50 to 0.800. (b) When the KOH eluent is used, the water dip is
much smaller. See Figure 4.24 for the DI water chromatogram using KOH eluent. Note that the
scale on the Y-axis is -0.300 to 0.800. Note also that the DI water chromatogram has a carbonate
peak positioned after chloride and before sulfate. The height of the carbonate peak will depend
on water quality, the age of the water and ambient room carbon dioxide levels. DI water
chromatograms must be free of the anions of interest before starting the analytical run.
8) Start the run. Run calibration standards first. The injection should start with the highest
concentration standard followed by decreasing concentrations.
9)
Run a low QCS directly after completing the low calibration curve and a high QCS after
completing the high calibration curve. Inject a QCS (randomly selected, high or low) every
ten samples thereafter. Plot the QCS results on control charts.
10) Calibrate every 30 to 50 samples.
11) Following the run, check all calibration curves and QCS results before reporting, collating or
tabulating sample results. Use only the peak area, not the peak height, for calculating results.
12) Examine each chromatogram individually for correct shape and integration. The baseline must
not drift up or down and must not be bumpy. All carbonate chromatograms should have a water
dip and all hydroxyl chromatograms should have a carbonate peak. The peaks should all have a
typical Gaussian shape and show good separation from each other. Comment on all anomalies
and flag data accordingly. Repeat samples that have drifting or bumpy baselines after resolving
the cause. See Troubleshooting, below.
Note that IC software uses peak ‘windows’. The software expects the peak for each analyte to
elute in a certain window of time. Peaks that fall outside this window will not be integrated. Also,
very large peaks that fill the window may not be recognized and thus produce a zero result.
Make note of these exceptions and repeat the analysis.
13) Calculate the final results against the appropriate calibration curve. Use the correct decimal
places. Apply detection limit notations as needed. Mark all samples that exceed the upper
calibration ranges. Dilute these samples and repeat the analysis. Account for any missing
samples and ensure contamination codes are applied as needed.
14) Export the data from the IC system and archive all parameters associated with the analysis,
including calibration data, integration data, and instrument audit trails. Audit trails include
instrument parameters (e.g. pump pressure) that may be useful in diagnosing a problem, such as
a chromatogram with a drifting baseline. It may be necessary to repeat the analysis at the point
where the problem began.